Apple’s own validation reports a mean VO₂max error of +1.4 ± 4.7 mL/kg/min, against a projected reference
What does Apple say about Cardio Fitness accuracy?
The finding
Validity error was 1.2 ± 4.4 mL/kg/min in the design set and 1.4 ± 4.7 in the validation set. Test-retest reliability ICC was 0.86 (0.80–0.90).
- Authors
- Apple Inc.
- Published
- Using Apple Watch to Estimate Cardio Fitness with VO₂ max (white paper) · 2021
- Evidence
- Read in the primary source; numbers checked.
- Source
- www.apple.com/healthcare/docs/site/Using_Apple_Watch_to_Estimate_Cardio_Fitness_with_VO2_max.pdf
How it was studied
Manufacturer study on Apple Watch Series 4: design set of 534, validation set of 221 (mean age 55 ± 17; 38% over 65). The reference was a linear projection of submaximal heart rate and VO₂ to age-predicted maximum heart rate, not a measured maximum.
The numbers
Estimates range from 14 to 60 mL/kg/min and need an outdoor walk, run or hike on a grade under 5%. Apple’s Low Cardio Fitness alerts use FRIEND’s lowest quintile for ages 20–59, and 18 (men) or 15 (women) mL/kg/min from age 60.
What it does not show
Written by the vendor, and the reference is itself an estimate rather than a measured VO₂max.
What Longstanding does with it
Apple and the app both read VO₂max against measured FRIEND norms, so the two should broadly agree. A watch value is still an estimate, and the app labels it as one and treats it as a trend.